Test Instrument per BGV A3 (previously VBG 4) and MPG (medial product law)
Universal tester for testing the electrical safety of portable electrical equipment
used in commercial and medical applications.
optional
In order to evaluate electrical safety, tests are executed for
protective conductor connections, insulation resistance and leakage current
(differential current, equivalent leakage current, device leakage current,
patient leakage current and contact current).
Technical Safety Testing for
- Electrical equipment in accordance with DIN VDE 0701, part 1 edition 2000-9 and part 200
- Information technology devices and equipment in accordance with DIN VDE 0701, part 240
- Periodic testing per DIN VDE 0702 (BGV A3)
- Electrical medical devices in accordance with DIN VDE 0751
Expanded functions
- Automatic recognition of safety class (I or II) and mains connection errors
- Menu-driven test sequences: fully automated or manual
- Patient leakage current: AC / DC components measured separately
- Equivalent leakage current with 230 V test voltage
- Convenient memory and report generating functions
- Data interface for PC, printer and barcode scanner
- Compact design, very lightweight
- DKD calibration certificate (optional)
Standard Models (include all basic features 00 [B00-K00]:
| Article No. |
Include Additional Features |
Feature |
| M7030-V001 |
Test sockets |
F01 |
| M7030-V002 |
Test sequence for hospital beds |
D07 |
| M7030-V003 |
with printer modul PSI |
E01 |
| M7030-V005 |
Test sockets and printer modul PSI |
F01, E01 |
| |
|
|
M701C, ...
-> more... |
Models for out of Germany with 10A protective
conductor test current but without medical testing per DIN VDE 0751 |
Software options:
| SE-L.med |
Foreign language user interface |
| DB-med |
DBmed database |
| SECU-dd |
Direct printing |
| SK5 |
Remote control Auto Measuring |
| SECU-Up |
Software upgrade for older series SII, SIII instr. |
Automated and systematic testing
|