- Application: universal current measurements
- Connector: 4 mm safety banana plugs
- Output: voltage e.g. current
Current Transformer Clamps -> Current Output mA/A
Current Sensor Clamps -> Voltage Output mV/A
Current Clamp Meters with Display - click here
Current
Probes and Current Sensors - for Power Analyzers - here
AC-/DC Current Sensors
- Voltage Output |
Type |
Nom.Current
A DC ;AC |
CAT
@..V |
Cross
Ø mm |
Sensitivity |
Frequency
Range |
Intrinsic Error
±(% v.M.+mA) |
Article
No. |
with
Battery
30h
30h
30h |
CP30 |
0.. 30 A
0.. 30 A peak |
III
300 |
25 |
100 mV/A |
DC..20kHz
(-1dB) |
1% *0,002A |
Z201B |
| CP330 |
0.. 30 A
0.. 300 A |
III
300 |
25 |
10 mV/A
1 mV/A |
DC..20kHz
(-3dB) |
1% +0,05A
1% +0,1A |
Z202B |
| CP1100 |
0..100 A
0..1000 A |
III
300 |
32 |
10 mV/A
1 mV/A |
DC..20kHz
(-1dB) |
1% +0,05A
1% *0,1A |
Z203B |
with
Batt.
50h |
Z13B |
0,5..60/600A
0,2..40/400A~ |
III
300 |
50 |
10 mV/A
1 mV/A |
DC..65Hz
..10kHz |
1,5%
2% |
Z213B |
AC Current Sensors
- Voltage Output |
Type |
Nom.Current
A AC |
CAT
@..V |
Cross
Ø mm. |
Sensitivity |
Frequency
Range |
Intrinsic Error
±(% v.M.+mA) |
Article
No. |

|
WZ11B |
0,5-20
5-200 |
III
600 |
20 |
100mV/A
10mV/A |
30..48..65
..500Hz |
1.. 3% |
Z208B |
| WZ12B |
0,01-100 |
III
300 |
15 |
0,1 mV/mA |
45..65
..500Hz |
1,5% +1 mA |
Z219B |
| WZ12C |
0,001-15
1-150 |
III
300 |
15 |
1mV/mA
1mV/A |
48..65
..500Hz |
3% +0,15mA
2%+0,1mA |
Z219C |
| WZ12E |
0,2-150 |
|
15 |
10 mV/A |
|
2% +10 mA |
Z823D |
| WZ12F |
0,02-15 |
|
15 |
100 mV/A |
|
2% +1mA |
Z823E |
| Clip 0100S |
0,001-15
1-150 |
|
15 |
1mV/mA
1mV/A |
|
3% +0,15mA
2%+0,1mA |
Z501E |
 |
Z3512A |
0,001-1/10/
/100/1000
|
III
600 |
52 |
1000/100
/10/1 mV/A |
10..48..65
..3kHz |
0,2%.. 0,7% |
Z225A |
|
Z821B |
1-3000 |
|
64 |
0,33mV/A |
|
0,5% +0,5 mV |
Z821B |
 |
METRAFLEX 300M |
3 A
30 A
300 A |
III @
1000
IV @
600 |
50 |
1 V/A
100 mV/A
10 mV/A |
20 Hz
...100 kHz |
1% +0,2 A
1% +0,2 A
1% +1 A |
Z207M |
 |
METRAFLEX P300 |
3 A
30 A
300 A |
III
1000
IV
600 |
140 |
1 V/A
100 mV/A
10 mV/A |
20 Hz
...4 kHz |
1% +0,01 A
1% +0,1 A
1% +1 A |
Z502E |
 |
METRAFLEX 3000 |
30 A
300 A
3000 A |
III @
1000
IV @
600 |
200 |
100mV/A
10 mV/A
1 mV/A |
10 Hz
... 20 kHz |
1% +0,1 A
1% +0,1 A
1% + 1 A |
Z207E |
AC Current
Transformers
- Current Output |
Type |
Nom.Current
A AC |
CAT
@..V |
Cross
Ø mm |
Sensitivity
mV/A |
Frequency
Range |
Intrinsic Error
±(%v.VAusg+mV) |
Article
No. |
|
WZ12A |
15.. 180 |
III
300 |
15 |
1mA/A |
45..65
..400Hz |
3% |
Z219A |
| WZ12D |
0,03-150 |
III
300 |
15 |
1mA/A |
45..65
..500Hz |
2,5% +1mA |
Z219D |

|
WZ11A |
1.. 200 |
III
600 |
20 |
1mA/A |
48..65
..400Hz |
1.. 3% |
Z208A |
| Z3511 |
4-500 |
III
600 |
30x63 |
1mA/A |
45..65
..1kHz |
3% +0,4 A |
GTZ3511000R0001 |
  |
Z3512 |
0,5-1000 |
III
600 |
52 |
1mA/A |
30..45..65
..5kHz |
0,5%.. 0,7% |
GTZ3512000R0001 |
| Z3514 |
1-2000 |
III
600 |
64 |
1mA/A |
30..45..65
..5kHz |
0,5% |